Effect of Leak and Breathing Pattern on the Accuracy of Tidal Volume Estimation – A Bench Study

Luján M, Sogo A, Pomares X, et al
Citation: Respir Care. 2013 May;58(5):770-7,

Abstract

BACKGROUND:

New home ventilators are able to provide clinicians data of interest through built-in software. Monitoring of tidal volume (VT) is a key point in the assessment of the efficacy of home mechanical ventilation.

OBJECTIVE:

To assess the reliability of the VT provided by 5 ventilators in a bench test.

METHODS:

Five commercial ventilators from 4 different manufacturers were tested in pressure support mode with the help of a breathing simulator under different conditions of mechanical respiratory pattern, inflation pressure, and intentional leakage. Values provided by the built-in software of each ventilator were compared breath to breath with the VT monitored through an external pneumotachograph. Ten breaths for each condition were compared for every tested situation.

RESULTS:

All tested ventilators underestimated VT (ranges of -21.7 mL to -83.5 mL, which corresponded to -3.6% to -14.7% of the externally measured VT). A direct relationship between leak and underestimation was found in 4 ventilators, with higher underestimations of the VT when the leakage increased, ranging between -2.27% and -5.42% for each 10 L/min increase in the leakage. A ventilator that included an algorithm that computes the pressure loss through the tube as a function of the flow exiting the ventilator had the minimal effect of leaks on the estimation of VT (0.3%). In 3 ventilators the underestimation was also influenced by mechanical pattern (lower underestimation with restrictive, and higher with obstructive).

CONCLUSIONS:

The inclusion of algorithms that calculate the pressure loss as a function of the flow exiting the ventilator in commercial models may increase the reliability of VT estimation.

Related Content
Home Mechanical Ventilation Through Mask-monitoring Leakage and Nocturnal S...
Fernandez Alvarez R, Rubinos Cuadrado G, Rodriguez Jerez F, et al
Product Query